Implementation of High-Temperature Superconducting Tapes RF Coils for 3T MRI System
Journal
IEEE Transactions on Magnetics
Pages
7413-7416
Date Issued
2005-09
Author(s)
Abstract
One way to reduce receiving coil noise in MRI scans is using non-resistive high-temperature superconducting (HTS) coils [1]. They show advantages of much lower cost and easier fabrication over HTS thin iilm coils. In this work, we built a 200mm in diameter Bi2Sr2Ca2Cu3Ox(Bi-2223) tape HTS RF coil and demonstrated that the SNR of using the HTS tape coil was 2.22 folds higher than that of the traditional copper coil for a phantom MR study. Test results were in agreement with predictions, and the error of predicted SNR gains and measured SNR gains is about 0.9%. The HTS coil can be expected to generate higher SNR gain after optimization. In the future, in-vivo experiments will be conducted to farther test the capability of the HTS tape coil. Further applications functional MRI is under investigation to test the power of this HTSC system in our 3T system.
SDGs
Type
conference paper