Multi-User Surgical Navigation Platform Based on Mixed Reality
Journal
IEEE Access
Journal Volume
12
Start Page
162522
End Page
162535
ISSN
2169-3536
Date Issued
2024
Author(s)
Abstract
Extended Reality (XR) technology enables visualization of patient data during surgery, with Augmented Reality (AR) systems offering 2D or 3D navigation. However, current systems face limitations such as lack of depth in 2D navigation, restricted single-user functionality in 3D navigation, and limited interoperability between tracking markers and navigation platforms. Additionally, users cannot interact with virtual objects, reducing operational efficiency and intuitiveness. This study addresses these issues by proposing a multi-user Mixed Reality (MR) guidance platform that provides interactive, real-time feedback and enhances collaboration. The platform maps 3D medical models onto real-world objects using 2D marker calibration, supporting focused ultrasound for transcranial brain therapy. Integration with BrainLab and HoloLens 2 achieved a single-user positioning accuracy of 0.576±0.294±with multi-user alignment errors averaging 2.718±1.335 mm. The system maintained a frame rate of 50-60 FPS, demonstrating robust performance and precision.
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Type