Non-invasive and time-dependent blood-sugar monitoring via breath-derived CO2 correlation using gas chromatograph with a milli-whistle gas analyzer
Journal
Analytical Sciences
Journal Volume
36
Journal Issue
6
Pages
739-743
Date Issued
2020
Author(s)
Lin, C.-H.
Wu, L.-X.
Chen, K.-H.
Lo, H.-F.
KING-CHUEN LIN
Kasai, T.
Chen, C.-C.
Shih, C.-H.
Manzano, M.C.
Santos, G.N.
Manzano, E.
Yu, D.E.
Abstract
A clear and positive correlation between the CO2 concentration and the blood-sugar level has been observed via a noninvasive and time-dependent monitoring of CO2 concentration from human breath, which is carried out by using a homemade gas chromatography (GC)/milli-whistle compact analyzer. The time-dependent sampling of the CO2 concentration correlated between 5.0 to 5.6% (1% = 104 ppm) in accordance with blood-sugar level variations of 80 to 110 mg/dL. The analytical method results in a rapid, continuous and non-invasive determination of blood-sugar level via measurement of the CO2 concentration exhaled from the lungs.