Development of micromachined electrochemical sensor and portable meter system
Resource
Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
Journal
20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society
Pages
-
Date Issued
1998-11
Date
1998-11
Author(s)
DOI
N/A
Abstract
An integrated microsensor has been designed and developed by using standard semiconductor technology and post-processing by micromachined procedures. The microsensor is implemented by using the UMC 0.5 /spl mu/ CMOS technology with a die size of 1.0 mm/spl times/1.0 mm. The microsensor consists of a two-electrode electrochemical cell, current-to-voltage converter and operational amplifier. The electrode area is 600 /spl mu//spl times/800 /spl mu/. The complete system contains a microsensor strip, a single chip microprocessor (80C196KC) and a LCD for portable instrument prototype. The system is capable of both amperometric and cyclic voltametric measurement for continuous monitoring and detail analysis of electrode response.