Coronary stent strut avulsion in aorto-ostial in-stent restenosis: Potential complication after cutting balloon angioplasty
Journal
Catheterization and Cardiovascular Interventions
Journal Volume
56
Journal Issue
2
Pages
215-219
Date Issued
2002
Author(s)
Abstract
We report a case of stent strut avulsion by the cutting balloon during the withdrawal of the deflated balloon catheter in aorto-ostial in-stent restenosis, which was managed successfully by another stent. The proposed mechanisms and recommendations to avoid this rare complication are provided. ? 2002 Wiley-Liss, Inc.
Subjects
Aorto-ostial lesion; Cutting balloon angioplasty; In-stent restenosis; Stent; Stent strut disruption
SDGs
Other Subjects
aged; article; case report; coronary artery disease; coronary stent; female; fluoroscopy; human; percutaneous transluminal angioplasty; postoperative complication; restenosis; Aged; Angioplasty, Transluminal, Percutaneous Coronary; Coronary Restenosis; Device Removal; Female; Humans; Stents