Evaluation of Bruker Biotyper and Vitek MS for the identification of Candida tropicalis on different solid culture media
Journal
Journal of Microbiology, Immunology and Infection
Journal Volume
52
Journal Issue
4
Pages
604-611
Date Issued
2019
Author(s)
Abstract
BACKGROUND: The aim of this study was to investigate the performance of the Bruker Biotyper matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) and Vitek MS systems for identification of genetically-confirmed blood isolates of Candida tropicalis that had been grown on several types of culture media commonly used for primary fungal isolation. METHODS: Isolates included 105 from the National China Hospital Invasive Fungal Surveillance Net program (CHIF-NET) and 120 from National Taiwan University Hospital (NTUH). Culture media tested for CHIF-NET isolates included trypticase soy agar supplemented with 5% sheep blood (BAP), Sabouraud dextrose agar (SDA-C), CHROMagar, China blue agar (CBA), chocolate agar supplemented with vancomycin (CAP-VA), and MacConkey agar (MAC). Culture media used for NTUH isolates included BAP, SDA, CHROMagar, eosin methylene blue (EMB), inhibitory mold agar (IMA), Mycosel agar, and cornmeal agar (CMA). RESULTS: The Bruker Biotyper correctly identified all CHIF-NET isolates to the species level on all six agar media tested and correctly identified the majority of NTUH isolates with the exception of isolates grown on SDA (85.8%) and CMA (52.5%). The Vitek MS system correctly identified all CHIF-NET isolates to the species level with the exception of isolates grown on CHROMagar (84.8%), and correctly identified the majority of NTUH isolates with the exception of isolates grown on SDA (51.7%), Mycosel agar (57.5%), and CMA (9.2%) for NTUH isolates. CONCLUSION: Clinical microbiologists should be aware that different culture media can affect the performance of the Bruker Biotyper MALDI-TOF MS and Vitek MS systems in identifying C. tropicalis.
SDGs
Type
journal article