Calibration strategy of the JUNO experiment
Journal
Journal of High Energy Physics
Journal Volume
2021
Journal Issue
3
Date Issued
2021
Author(s)
Abulencia, A. et al. (CDF Collaboration)
Abstract
We present the calibration strategy for the 20 kton liquid scintillator central detector of the Jiangmen Underground Neutrino Observatory (JUNO). By utilizing a comprehensive multiple-source and multiple-positional calibration program, in combination with a novel dual calorimetry technique exploiting two independent photosensors and readout systems, we demonstrate that the JUNO central detector can achieve a better than 1% energy linearity and a 3% effective energy resolution, required by the neutrino mass ordering determination. [Figure not available: see fulltext.] ? 2021, The Author(s).