Repetition rate multipliers for efficient implementation of multiphoton and pump-probe fluorescence microscopy
Journal
Optik
Journal Volume
114
Journal Issue
8
Pages
337-342
Date Issued
2003
Author(s)
Abstract
With the advances in pulsed laser systems, microscopic imaging techniques such, as multiphoton and pump-probe fluorescence microscopy have developed into effective tools for investigating intensity and time-resolved phenomena inside biological systems. However, pulsed lasers used in these techniques usually are commercial systems with repetition frequencies of around 80 MHz. While these systems have proven to be adequate for multiphoton and pump-probe microscopic imaging applications, the temporal separation of the laser pulse train (around 12.5 ns) is long compared to the fluorescence lifetimes of many common fluorescence species. In this work, we present the designs of repetition rate multipliers based on passive optical components that can be used to increase the efficiency in multiphoton and pump-probe fluorescence microscopy. Depending on the lifetime of fluorescence molecules under investigation, the passive repetition rate multiplier can increase the duty cycle of multiphoton or pump-probe microscopy up to fourfold.