Sensors for the CDF Run2b silicon detector
Journal
IEEE Transactions on Nuclear Science
Journal Volume
51
Journal Issue
4 I
Pages
1546-1554
Date Issued
2004
Author(s)
Akimoto, H. et al. (CDF Collaboration)
Abstract
We describe the characteristics of silicon microstrip sensors fabricated by Hamamatsu Photonics for the CDF Run 2b silicon detector. A total of 953 sensors, including 117 prototype sensors, have been produced and tested. Five sensors were irradiated with neutrons up to 1.4 /spl times/10/sup 14/ n/cm/sup 2/ as a part of the sensor quality assurance program. The electrical and mechanical characteristics are found to be superior in all aspects and fulfill our specifications. We comment on charge-up susceptibility of the sensors that employ a crystal structure.
SDGs
Type
conference paper