Glucose biosensor of ruthenium-doped TiO2 sensing electrode by co-sputtering system.
Journal
Microelectronics Reliability
Journal Volume
50
Journal Issue
5
Pages
753-756
Date Issued
2010
Author(s)
Abstract
In this study, the ruthenium-doped titanium dioxide (TiO2:Ru) sensing film is fabricated by co-sputtering system. The TiO2:Ru sensing film is treated by thermal annealing process at 600 °C for 1 h in atmosphere. We present a TiO2:Ru-based pH sensor which can obtain the higher sensing value at 64.58 mV/pH for the super-Nernstian response. The enzyme composite solution is dropped on the TiO2:Ru sensing film as a glucose biosensor. In the concentration ranging from 100 to 500 (mg/dL) of glucose solution, the sensitivity and linearity of TiO2:Ru sensing film after thermal annealing are 0.320 mV(mg/dL)-1 and 0.995, respectively, which is better than non-annealing TiO2:Ru sensing film. In this study, we can describe a novel doping method to prepare the TiO2:Ru sensing film and the glucose biosensors. © 2010 Elsevier Ltd. All rights reserved.