Microvascular imaging using swept-source optical coherence tomography with single-channel acquisition
Journal
Appl. Phys. Express
Journal Volume
4
Pages
097001-097003
Date Issued
2011
Author(s)
M-T Tsai
T-D Chi
H-L Liu
F-Y Chang
C-H Yang
C-K Lee
C-C Yang
M-T Tsai
T-D Chi
H-L Liu
F-Y Chang
C-H Yang
C-K Lee
C-C Yang
Abstract
This study proposes a new approach to improve the phase stability for swept-source optical coherence tomography (SS-OCT) with single-channel acquisition. This approach can improve the phase instability due to the A-scan trigger jitter from the swept source, or the asynchronization between the A-scan trigger and high-speed digitizer, which enables the visualization of vascular structures by SS-OCT. Aside from reducing the phase noise of the OCT system, only one channel is required for the A-scan trigger, data storage, and wavelength calibration by implementing the proposed approach. Finally, human skin was scanned in vivo to demonstrate the vascular images.