標題: | Patient-specific instrumentation improves functional kinematics of minimally-invasive total knee replacements as revealed by computerized 3D fluoroscopy | 作者: | Shih, K.-S. Lin, C.-C. Lu, H.-L. Fu, Y.-C. Lin, C.-K. Li, S.-Y. Lu, T.-W. TUNG-WU LU |
關鍵字: | Isolated knee exercise; Joint biomechanics; Surface kinematics; Three-dimensional fluoroscopy | 公開日期: | 2020 | 卷: | 188 | 來源出版物: | Computer Methods and Programs in Biomedicine | 摘要: | Background and objectives: Minimally-invasive total knee arthroplasty (MIS-TKA) has demonstrated very good short-term success, but its mid- to long-term results remain inconclusive. The success may be related to the tradeoff between a small incision and accurate positioning of the implant components. Patient-specific instrumentation (PSI) aims to improve the accuracy in restoring the knee axis and the clinical outcomes for MIS-TKA, but the results are yet to be confirmed by accurate assessment during functional activities. The purpose of the current study was to measure and compare the in vivo three-dimensional (3D) rigid-body and surface kinematics of MIS-TKA implanted with and without PSI during isolated knee active flexion/extension and sit-to-stand using state-of-the-art 3D model-based fluoroscopy technology. Methods: Ten patients treated for advanced medial knee osteoarthritis by MIS-TKA without PSI (non-PSI group) and nine with PSI (PSI group) participated in the current study. Each subject performed non-weight-bearing knee flexion/extension and sit-to-stand tasks while the motion of the prosthetic knee was under bi-plane fluoroscopy surveillance. The computer models of each of the knee prosthesis components were registered to the measured fluoroscopy images for each time frame via a novel validated 3D fluoroscopy method. Non-parametric 1-tailed Mann-Whitney tests were performed to detect the differences in the joint and surface kinematic variables every 10° of knee flexion between the non-PSI and PSI groups. The 1-tailed significance level was at α = 0.05. Results: The PSI group showed clear, coupled flexion/internal rotation during activities, while the non-PSI group remained roughly at an externally rotated position with slight internal rotations. The coupled rotation in the PSI group was accompanied by an anterior displacement of the medial contact and a posterior displacement of the lateral contact, which was different from the screw-home mechanism. Neither of the two groups showed the normal roll-back phenomenon, i.e., posterior translation of the femur relative to the tibia during knee flexion. Conclusions: With the state-of-the-art 3D fluoroscopy method, differences in both the rigid-body and surface kinematics of the prosthetic knees between MIS-TKA with and without PSI were identified. Patients with PSI demonstrated significant positive effects on the reconstructed rigid-body kinematics of the knee, showing clearer coupled flexion/internal rotations – an important kinematic characteristic in healthy knees – than those without PSI during activities with or without weight-bearing. However, none of them showed normal contact patterns. The current findings will be helpful for surgical instrument design, as well as for surgical decision-making in MIS total knee arthroplasty. ? 2019 |
URI: | https://www.scopus.com/inward/record.url?eid=2-s2.0-85076260080&partnerID=40&md5=3a240f180b175b5dc580f793e08bd4a9 https://scholars.lib.ntu.edu.tw/handle/123456789/546999 |
DOI: | 10.1016/j.cmpb.2019.105250 | SDG/關鍵字: | 3D modeling; Arthroplasty; Bone; Decision making; Fluorescent screens; Implants (surgical); Joints (anatomy); Kinematics; Knee prostheses; Orthopedics; Patient monitoring; Rigid structures; Rotation; Surgical equipment; Functional activities; Isolated knee exercise; Joint biomechanics; Kinematic characteristics; Rigid body kinematics; Threedimensional (3-d); Total knee arthroplasty; Total knee replacement; Physiological models; aged; Article; clinical article; clinical outcome; computer model; female; femur; fluoroscopy; human; joint mobility; kinematics; knee function; knee osteoarthritis; male; minimally invasive surgery; three dimensional imaging; tibia; total knee arthroplasty; biomechanics; diagnostic imaging; image processing; joint characteristics and functions; knee; knee osteoarthritis; knee prosthesis; knee replacement; middle aged; pathophysiology; prosthesis design; reproducibility; three-dimensional imaging; weight bearing; Aged; Arthroplasty, Replacement, Knee; Biomechanical Phenomena; Female; Femur; Fluoroscopy; Humans; Image Processing, Computer-Assisted; Imaging, Three-Dimensional; Knee; Knee Joint; Knee Prosthesis; Male; Middle Aged; Minimally Invasive Surgical Procedures; Osteoarthritis, Knee; Prosthesis Design; Range of Motion, Articular; Reproducibility of Results; Weight-Bearing |
顯示於: | 醫學工程學研究所 |
在 IR 系統中的文件,除了特別指名其著作權條款之外,均受到著作權保護,並且保留所有的權利。