Title: | Fracture of a perfusion balloon catheter during coronary angioplasty | Authors: | Kuan, Peiliang Ko, Yu‐Lin CHENG-HSIU TSAI JIUNN-LEE LIN |
Keywords: | arteriotomy | perfusion balloon catheter | PTCA complication | Issue Date: | May-1994 | Journal Volume: | 32 | Journal Issue: | 1 | Start page/Pages: | 33 | Source: | Catheterization and cardiovascular diagnosis | Abstract: | In a 74-year-old patient with increasing angina, an attempt was made to dilate an ostial stenosis in the first diagonal artery. The initial attempt with a long balloon catheter was unsuccessful. Because of chest pain associated with hypotension during balloon inflations, a second attempt with a perfusion balloon catheter resulted in fracture of this catheter's tip (14 mm long) in the artery. Bypass graft surgery was performed and the retained fragment was easily removed by arteriotomy. The patient made an uneventful recovery. |
URI: | https://www.scopus.com/record/display.uri?eid=2-s2.0-0028197395&doi=10.1002%2fccd.1810320107&origin=inward&txGid=1d272fc3d0016c28c797fda594eac65b https://scholars.lib.ntu.edu.tw/handle/123456789/634514 |
ISSN: | 0098-6569 | DOI: | 10.1002/ccd.1810320107 |
Appears in Collections: | 醫學系 |