https://scholars.lib.ntu.edu.tw/handle/123456789/637761
標題: | Using a short-term parameter of heart rate variability to distinguish awake from isoflurane anesthetic states | 作者: | HUI-HSUN HUANG Lee, Yi-Hui Chan, Hsiao-Lung YONG-PING WANG CHI-HSIANG HUANG SHOU-ZEN FAN |
關鍵字: | Anesthetic depth | Biomedical signal processing | Heart rate variability | Isoflurane | Prediction probability | 公開日期: | 十月-2008 | 卷: | 46 | 期: | 10 | 來源出版物: | Medical & biological engineering & computing | 摘要: | The measurement of anesthetic depth is important in anesthesiology. Although heart rate variability (HRV) is profoundly affected by general anesthesia, it has not yet been commonly used in this field. One of the reasons is the lack of suitable parameters of HRV for short-term observations. In this study, we designed a time domain parameter of HRV named the similarity index. It was based on observing the trend of the distribution of instantaneous heart rates as time moved. Taking epochs of ECG data as short as 64 s can derive the index. We observed the values of this index of 30 patients when they were awake and under isoflurane anesthesia. The values had very little overlapping between the two states and the prediction probability to distinguish the two states was 0.91. We suggest that HRV, if suitably treated, can play more roles in the monitoring of anesthetic depth. |
URI: | https://scholars.lib.ntu.edu.tw/handle/123456789/637761 | ISSN: | 01400118 | DOI: | 10.1007/s11517-008-0342-y |
顯示於: | 醫學系 |
在 IR 系統中的文件,除了特別指名其著作權條款之外,均受到著作權保護,並且保留所有的權利。